Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 16 de 16
Filtrar
1.
Pacing Clin Electrophysiol ; 36(12): 1539-49, 2013 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-24117856

RESUMO

BACKGROUND: Although several studies have demonstrated deleterious consequences of chronic right ventricular (RV) pacing on ventricular function and synchronicity, its effects on health-related quality of life (HRQoL) and functional exercise capacity remain uncertain. We aimed to evaluate the effect of RV pacing on HRQoL and functional capacity of children and young adults with congenital complete atrioventricular block (CCAVB). METHODS: We included 66 consecutive subjects with RV cardiac pacemaker due to CCAVB and under clinical follow-up for more than 1 year. Multidimensional HRQoL scores were evaluated by Short Form-36 Health Survey (SF-36) and Child Health Questionnaire-Parent Form 50 (CHQ-PF50). Functional capacity was tested by a 6-minute walk distance test (6MWDT). Association analysis was performed to examine if any demographic and clinical characteristics were associated with lower HRQoL scores and shorter distances at the 6MWDT. RESULTS: Domains presenting lower HRQoL scores were vitality (64.0 ± 17.3), mental health (67.3 ± 8.0), role emotional (69.7 ± 35.8) in the SF-36 questionnaire; general health perceptions (61.3 ± 8.3), general behavior (61.9 ± 15.6), parental impact-emotional (67.7 ± 28.7) in the CHQ-PF50. Female gender (P = 0.009), left ventricular ejection fraction lower than 55% (P = 0.013), cardiovascular drugs (P = 0.003) were significantly associated with lower HRQoL scores. Average distance traveled during the 6MWDT was 539.8 ± 82.9 m. The 6MWDT showed significant association with age (P = 0.006) and cardiovascular drugs (P = 0.024). CONCLUSIONS: Chronic RV pacing did not affect the HRQoL and physical capacity of pediatrics and young subjects. Female gender, ventricular function, and cardiovascular drugs were associated with lower HRQoL scores. Older subjects walked shorter distances in the 6MWDT, as well as subjects who were taking cardiovascular drugs.


Assuntos
Bloqueio Atrioventricular/congênito , Bloqueio Atrioventricular/psicologia , Estimulação Cardíaca Artificial/métodos , Estimulação Cardíaca Artificial/psicologia , Qualidade de Vida/psicologia , Recuperação de Função Fisiológica , Adolescente , Distribuição por Idade , Bloqueio Atrioventricular/prevenção & controle , Criança , Pré-Escolar , Feminino , Ventrículos do Coração , Humanos , Recém-Nascido , Estudos Longitudinais , Masculino , Distribuição por Sexo , Resultado do Tratamento , Adulto Jovem
2.
Acta Cir Bras ; 28(7): 543-50, 2013 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-23842937

RESUMO

PURPOSE: To assess the effectiveness of bipolar epicardial atrial pacing using an active fixation bipolar endocardial lead implanted on the atrial surface in an experimental model. METHODS: A total of ten Large White adult pigs underwent pacemaker implantation under general anesthesia. Atrial pacing and sensing parameters were obtained at the procedure, immediate postoperative period and on the 7th and the 30th postoperative in unipolar and bipolar configurations. RESULTS: All procedures were successfully performed. There were no perioperative complications and no early deaths. Atrial pacing and sensing parameters for both unipolar and bipolar modes remained stable throughout the study. We observed a progressive increase in atrial thresholds, ranging from 0.49 ± 0.35 (at implantation) to 1.86 ± 1.31 volts (30th postoperative day), in unipolar mode. Atrial impedance measurements decreased slightly over time, ranging from 486.80 ± 126.35 Ohms (at implantation) to 385.0 ± 80.52 Ohms (30th postoperative day). Atrial sensing measures remained stable from the immediate postoperative period until the end of the study. CONCLUSION: The bipolar active fixation endocardial lead implanted epicardially can provide stable conditions of pacing and sensing parameters throughout the postoperative follow-up.


Assuntos
Estimulação Cardíaca Artificial/métodos , Eletrodos Implantados , Desenho de Equipamento/métodos , Átrios do Coração/cirurgia , Marca-Passo Artificial , Esteroides/administração & dosagem , Animais , Átrios do Coração/fisiopatologia , Duração da Cirurgia , Período Pós-Operatório , Reprodutibilidade dos Testes , Propriedades de Superfície , Suínos , Fatores de Tempo , Resultado do Tratamento
3.
Acta cir. bras ; 28(7): 543-550, July 2013. ilus, graf, tab
Artigo em Inglês | LILACS | ID: lil-679088

RESUMO

PURPOSE: To assess the effectiveness of bipolar epicardial atrial pacing using an active fixation bipolar endocardial lead implanted on the atrial surface in an experimental model. METHODS: A total of ten Large White adult pigs underwent pacemaker implantation under general anesthesia. Atrial pacing and sensing parameters were obtained at the procedure, immediate postoperative period and on the 7th and the 30th postoperative in unipolar and bipolar configurations. RESULTS: All procedures were successfully performed. There were no perioperative complications and no early deaths. Atrial pacing and sensing parameters for both unipolar and bipolar modes remained stable throughout the study. We observed a progressive increase in atrial thresholds, ranging from 0.49 ± 0.35 (at implantation) to 1.86 ± 1.31 volts (30th postoperative day), in unipolar mode. Atrial impedance measurements decreased slightly over time, ranging from 486.80 ± 126.35 Ohms (at implantation) to 385.0 ± 80.52 Ohms (30th postoperative day). Atrial sensing measures remained stable from the immediate postoperative period until the end of the study. CONCLUSION: The bipolar active fixation endocardial lead implanted epicardially can provide stable conditions of pacing and sensing parameters throughout the postoperative follow-up.


Assuntos
Animais , Estimulação Cardíaca Artificial/métodos , Eletrodos Implantados , Desenho de Equipamento/métodos , Átrios do Coração/cirurgia , Marca-Passo Artificial , Esteroides/administração & dosagem , Átrios do Coração/fisiopatologia , Duração da Cirurgia , Período Pós-Operatório , Reprodutibilidade dos Testes , Propriedades de Superfície , Suínos , Fatores de Tempo , Resultado do Tratamento
4.
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm ; 22(3): 143-151, jul.-set. 2009. ilus, tab, graf
Artigo em Português | LILACS | ID: lil-538323

RESUMO

Objetivo: Comparar o comportamento clínico-evolutivo de portadores de CDI por taquicardia ventricular (TV) sincopal ou parada cardíaca (PC) por TV/FV. Método: 585 pacientes foram submetidos a implante de CDI entre jan/2000 e jul/2005 para prevenção secundária de morte súbita cardíaca (MSC). Foram selecionados 415 pacientes de um banco de dados prospectivo, distribuídos em dois grupos: G1, com TV sincopal (n=318) e G2, com PC por TV/FV (n=97). As variáveis analisadas foram: idade, sexo, fração de ejeção do ventrículo esquerdo(FEVE), classe funcional de insuficiência cardíaca (CF IC), medicamentos, terapias de choque apropriadas (TCA) pelo CDI e óbitos. Para análise estatística...


Assuntos
Humanos , Masculino , Feminino , Pessoa de Meia-Idade , Desfibriladores Implantáveis , Morte Súbita Cardíaca/prevenção & controle , Síncope/complicações , Taquicardia Ventricular/induzido quimicamente
5.
Rev. bras. anestesiol ; 58(4): 363-370, jul.-ago. 2008. ilus, tab
Artigo em Inglês, Português | LILACS | ID: lil-487164

RESUMO

JUSTIFICATIVA E OBJETIVO: Embora sejam os cães os animais mais utilizados em Cirurgia Experimental, nota-se crescente utilização de ovinos, bovinos e suínos como modelos para experimentação científica. Assim, faz-se necessário maior aprendizado de seu tratamento e padronizações básicas para os procedimentos cirúrgicos mais complexos em suínos. O objetivo foi avaliar a sedação e analgesia, obtidas pela injeção intramuscular de midazolam e cetamina e anestesia local com lidocaína a 2 por cento sem vasoconstritor, na realização de traqueostomia, dissecção de artéria e veia femorais. MÉTODO: A freqüência cardíaca e o reflexo córneo-palpebral foram avaliados logo no início da sedação, durante os procedimentos cirúrgicos descritos e após seu término. Foram utilizados oito suínos fornecidos por fazendas locais sem tratamento prévio da raça Large White, com peso de 35 a 42 quilos que receberam injeção intramuscular de 22 mg.kg-1 de cetamina e 0,3 mg.kg-1 de midazolam. RESULTADOS: A abordagem anestésica nesses animais, de difíceis intubação e acesso aos vasos profundos, provou ser segura para sedação e analgesia através do método utilizado: acesso intramuscular. CONCLUSÃO: Obteve-se adequado plano de anestesia para o procedimento proposto.


BACKGROUND AND OBJECTIVES: Although dogs are the animals used more often in Experimental Surgeries, the use of sheep, bovine, and swine as models for scientific experiments has been growing. Thus, further learning on the treatment and basic standardization of more complex surgeries in swine are necessary. The objective of this study was to evaluate sedation and analgesia, obtained by the intramuscular administration of midazolam and ketamine, and local anesthetics with 2 percent lidocaine without vasoconstrictor, in tracheostomy and dissection of femoral artery and vein. METHODS: The heart rate and corneal-palpebral reflex were evaluated at the beginning of sedation, during the surgical procedures, and at the end of those procedures. Eight Large White swine, without prior treatment, weighing 35 to 42 kg were provided by local farms. They received intramuscular injections of 22 mg.kg-1 of ketamine and 0.3 mg.kg-1 of midazolam. RESULTS: The anesthetic approach on those animals, who present both difficult intubation and access to deep vessels, proved to be safe for sedation and analgesia by the method chosen: intramuscular administration. CONCLUSIONS: An adequate plane of anesthesia for the proposed procedure was achieved.


JUSTIFICATIVA Y OBJETIVOS: Aunque los perros sean los animales más a menudo usados en Cirugía Experimental, se nota un aumento en la utilización de los ovinos, bovinos y porcinos como modelo para experimentación científica. De esa forma, se hace necesario aprender más con su tratamiento y estandarizaciones básicas para los procedimientos quirúrgicos más complejos en porcinos. El objetivo fue evaluar la sedación y la analgesia obtenidas por la inyección intramuscular de midazolán y la cetamina y anestesia local con lidocaína a un 2 por ciento sin vasoconstrictor, en la realización de traqueostomía, disección de arteria y vena femorales. MÉTODO: La frecuencia cardíaca y el reflejo córneo-palpebral fueron evaluados inmediatamente al inicio de la sedación, durante los procedimientos quirúrgicos descritos y después en su conclusión. Se utilizaron ocho porcinos que fueron suministrados por haciendas de la región sin tratamiento previo de la raza Large White, con peso de 35 a 42 kilos que recibieron inyección intramuscular de 22 mg.kg-1 de cetamina y 0.3 mg.kg-1 de midazolán. RESULTADOS: El abordaje anestésico en esos animales, de difícil intubación y acceso a los vasos profundos, demostró ser seguro para la sedación y analgesia a través del método utilizado: el acceso intramuscular. CONCLUSIÓN: Se obtuvo un adecuado plan de anestesia para el procedimiento propuesto.


Assuntos
Animais , Lidocaína/uso terapêutico , Midazolam/normas , Procedimentos Cirúrgicos Cardiovasculares/métodos , Suínos
6.
Arq Bras Cardiol ; 90(3): 160-6, 2008 Mar.
Artigo em Inglês, Português | MEDLINE | ID: mdl-18392394

RESUMO

BACKGROUND: The ideal programming of the implantable cardioverter defibrillator (ICD) shock energy should be at least 10J above the defibrillation threshold (DFT), requiring alternative techniques when the DFT is elevated. OBJECTIVE: To assess the clinical behavior of ICD patients with DFT>25J and the efficacy of the chosen therapy. METHODS: Patients who had undergone ICD implantation between Jan/00 and Aug/04 (prospective database) and presented intraoperative DFT>25J were selected. The analyzed variables were: clinical characteristics, LVEF, rescue of arrhythmic events from ICD and causes of deaths. RESULTS: among 476 patients, 16 (3.36%) presented DFT>25J. The mean age was 56.5 years, and 13 patients (81%) were men. According to the baseline cardiomyopathy, 09 patients had Chagas' disease, 04 had ischemic cardiomyopathy and 03 had idiopathic cardiomyopathy. Mean LVEF was 0.37 and amiodarone was used by 94% of the patients. Mean follow-up (FU) period was 25.3 months. DFT was higher than maximum energy shock (MES) in 2 patients and it was necessary to implant an additional shock electrode (array). It was programmed MES in ventricular fibrillation zone of ICD therapy in the other patients. In the FU, 03 patients had 67 successful appropriate shock therapies (AST). There were 05 noncardiac and 02 heart failure deaths. The patients who died showed higher DFT levels (p=0.044) without correlation with death because there wasn't unsuccessful AST. CONCLUSION: In this cohort of ICD patients, the occurrence of elevated DFT (>25J) was low, leading to alternative therapies. There was an association with severe ventricular dysfunction, although without correlation to the causes of death.


Assuntos
Arritmias Cardíacas/terapia , Desfibriladores Implantáveis , Cardioversão Elétrica , Amiodarona/uso terapêutico , Antiarrítmicos/uso terapêutico , Arritmias Cardíacas/tratamento farmacológico , Arritmias Cardíacas/mortalidade , Morte Súbita Cardíaca/etiologia , Morte Súbita Cardíaca/prevenção & controle , Desfibriladores Implantáveis/normas , Limiar Diferencial , Cardioversão Elétrica/normas , Técnicas Eletrofisiológicas Cardíacas , Feminino , Seguimentos , Humanos , Masculino , Pessoa de Meia-Idade , Estudos Prospectivos , Taquicardia Ventricular/terapia , Fibrilação Ventricular/terapia
7.
Arq. bras. cardiol ; 90(3): 177-184, mar. 2008. ilus, graf, tab
Artigo em Inglês, Português | LILACS | ID: lil-479618

RESUMO

FUNDAMENTO: A programação ideal da energia de choque do CDI deve ser pelo menos 10 J acima do limiar de desfibrilação (LDF), necessitando de técnicas alternativas quando o LDF é elevado. OBJETIVO: Avaliar o comportamento clínico dos portadores de CDI com LDF>25 J e a eficácia da terapêutica escolhida. MÉTODOS: Foram selecionados portadores de CDI, entre janeiro de 2000 e agosto de 2004 (banco de dados prospectivo), com LDF>25 J intra-operatório, e analisaram-se: características clínicas, FEVE, resgate de eventos arrítmicos pelo CDI e óbitos. RESULTADOS: dentre 476 pacientes, 16 (3,36 por cento) apresentaram LDF>25J. Idade média de 56,5 anos, sendo 13 pacientes (81 por cento) do sexo masculino. Quanto à cardiopatia de base 09 eram chagásicos, 04 isquêmicos e 03 com etiologia idiopática. A FEVE média dos pacientes foi 37 por cento e 94 por cento utilizavam amiodarona. O seguimento médio foi de 25,3 meses. Em 02 pacientes com LDF > Choque Máximo (CM), foi necessário implante de eletrodo de choque adicional (array), sendo mantido programação com CM em zona de FV (>182bpm) nos demais. Durante o seguimento 03 pacientes apresentaram 67 terapias de choque apropriadas (TCA) com sucesso. Ocorreram 07 óbitos sendo 5 por causas não cardíacas e 2 por insuficiência cardíaca avançada. Os pacientes que foram a óbito apresentaram níveis de LDF maiores (p=0,0446), entretanto sem relação com a causa dos mesmos tendo em vista que não ocorreram TCA sem sucesso. CONCLUSÃO: Nessa coorte de pacientes com CDI, a ocorrência de LDF elevado foi baixa, implicando terapêuticas alternativas. Houve associação com disfunção ventricular grave, entretanto sem correlação com as causas de óbito.


BACKGROUND: The ideal programming of the implantable cardioverter defibrillator (ICD) shock energy should be at least 10J above the defibrillation threshold (DFT), requiring alternative techniques when the DFT is elevated. OBJECTIVE: To assess the clinical behavior of ICD patients with DFT>25J and the efficacy of the chosen therapy. METHODS: Patients who had undergone ICD implantation between Jan/00 and Aug/04 (prospective database) and presented intraoperative DFT>25J were selected. The analyzed variables were: clinical characteristics, LVEF, rescue of arrhythmic events from ICD and causes of deaths. RESULTS: among 476 patients, 16 (3.36 percent) presented DFT>25J. The mean age was 56.5 years, and 13 patients (81 percent) were men. According to the baseline cardiomyopathy, 09 patients had Chagas' disease, 04 had ischemic cardiomyopathy and 03 had idiopathic cardiomyopathy. Mean LVEF was 0.37 and amiodarone was used by 94 percent of the patients. Mean follow-up (FU) period was 25.3 months. DFT was higher than maximum energy shock (MES) in 2 patients and it was necessary to implant an additional shock electrode (array). It was programmed MES in ventricular fibrillation zone of ICD therapy in the other patients. In the FU, 03 patients had 67 successful appropriate shock therapies (AST). There were 05 noncardiac and 02 heart failure deaths. The patients who died showed higher DFT levels (p=0.044) without correlation with death because there wasn't unsuccessful AST. CONCLUSION: In this cohort of ICD patients, the occurrence of elevated DFT (>25J) was low, leading to alternative therapies. There was an association with severe ventricular dysfunction, although without correlation to the causes of death.


Assuntos
Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Arritmias Cardíacas/terapia , Desfibriladores Implantáveis , Cardioversão Elétrica , Amiodarona/uso terapêutico , Antiarrítmicos/uso terapêutico , Arritmias Cardíacas/tratamento farmacológico , Arritmias Cardíacas/mortalidade , Limiar Diferencial , Morte Súbita Cardíaca/etiologia , Morte Súbita Cardíaca/prevenção & controle , Desfibriladores Implantáveis/normas , Técnicas Eletrofisiológicas Cardíacas , Cardioversão Elétrica/normas , Seguimentos , Estudos Prospectivos , Taquicardia Ventricular/terapia , Fibrilação Ventricular/terapia
8.
Rev Bras Anestesiol ; 58(4): 363-70, 2008.
Artigo em Inglês, Português | MEDLINE | ID: mdl-19378584

RESUMO

BACKGROUND AND OBJECTIVES: Although dogs are the animals used more often in Experimental Surgeries, the use of sheep, bovine, and swine as models for scientific experiments has been growing. Thus, further learning on the treatment and basic standardization of more complex surgeries in swine are necessary. The objective of this study was to evaluate sedation and analgesia, obtained by the intramuscular administration of midazolam and ketamine, and local anesthetics with 2% lidocaine without vasoconstrictor, in tracheostomy and dissection of femoral artery and vein. METHODS: The heart rate and corneal-palpebral reflex were evaluated at the beginning of sedation, during the surgical procedures, and at the end of those procedures. Eight Large White swine, without prior treatment, weighing 35 to 42 kg were provided by local farms. They received intramuscular injections of 22 mg.kg of ketamine and 0.3 mg.kg(-1) of midazolam. RESULTS: The anesthetic approach on those animals, who present both difficult intubation and access to deep vessels, proved to be safe for sedation and analgesia by the method chosen: intramuscular administration. CONCLUSIONS: An adequate plane of anesthesia for the proposed procedure was achieved.


Assuntos
Anestesia/normas , Procedimentos Cirúrgicos Cardiovasculares , Modelos Animais , Animais , Suínos
9.
Arq Bras Cardiol ; 88(3): 258-64, 2007 Mar.
Artigo em Inglês, Português | MEDLINE | ID: mdl-17533465

RESUMO

OBJECTIVES: To analyze the incidence and causes of ICD therapies in children and young adults and verify their impact on the quality of life (QoL). METHODS: From March/1977 to February/2006, 29 patients (15.7+/-5.4 years old) were submitted to ICD implants. Aborted cardiac arrest (41.5%), sustained ventricular tachycardia (27.6%) and primary prophylaxis of sudden cardiac death (30.9%) indicated device therapy. The number of therapies was evaluated by interviewing patients and by ICD diagnostic data. The SF-36 questionnaire was used to measure the QoL and the results were compared to healthy population. The expectative of freedom from ICD therapies were estimated by the Kaplan-Meier method. RESULTS: After 2.6+/-1.8 years follow-up, 8 (27.6%) patients received 141 appropriate ICD shocks due to ventricular tachycardia (6) or ventricular fibrillation (2), and 11 (37.9%) patients received 152 inappropriate ICD shocks due to supraventricular tachyarrhythmias (8) or oversensing (3). Expectative of freedom from appropriate shocks was 74.2+/-9.0% and 66.7+/-10.7% after one and three years, respectively. Compared to healthy population, QoL decreased in physical function (61.7+/-28.7), vitality (64.7+/-19.1), mental health (65.9+/-22.7) and role-emotional domains (66.7+/-38.5). All patients referred fear and concern related to ICD use. CONCLUSION: Despite the efficacy of ICD therapies, the high incidence of appropriate and inappropriate shocks interfered in patients QoL and adaptation to the device.


Assuntos
Arritmias Cardíacas , Desfibriladores Implantáveis/estatística & dados numéricos , Cardioversão Elétrica/estatística & dados numéricos , Qualidade de Vida , Adolescente , Adulto , Arritmias Cardíacas/etiologia , Arritmias Cardíacas/terapia , Criança , Pré-Escolar , Morte Súbita Cardíaca , Feminino , Parada Cardíaca/complicações , Parada Cardíaca/terapia , Humanos , Estimativa de Kaplan-Meier , Masculino , Inquéritos e Questionários , Taquicardia Supraventricular/complicações , Taquicardia Supraventricular/terapia
10.
Arq. bras. cardiol ; 88(3): 258-264, mar. 2007. tab, ilus
Artigo em Português | LILACS | ID: lil-451725

RESUMO

OBJETIVOS: Avaliar a incidência e a causa de choques de CDI em crianças e adolescentes e sua repercussão na qualidade de vida (QV). MÉTODOS: De março/1997 a fevereiro/2006, 29 pacientes (15,7±5,4 anos) foram submetidos a implante de CDI. Parada cardiorrespiratória recuperada (41,5 por cento), taquicardia ventricular sustentada (27,6 por cento) e profilaxia primária de morte súbita cardíaca (30,9 por cento) motivaram os implantes. O número de terapias foi avaliado por entrevista e pela telemetria dos CDI. A QV foi avaliada pela aplicação do questionário SF-36 e comparada à de indivíduos saudáveis. Empregou-se o método de Kaplan-Meier para análise da sobrevida livre de choques. RESULTADOS: Após 2,6±1,8 anos de seguimento, 8 (27,6 por cento) pacientes receberam 141 choques apropriados em razão de TV polimórfica (6) ou FV (2), e 11 (37,9 por cento) sofreram 152 choques inapropriados em razão de taquiarritmias supraventriculares (8) ou oversensing (3). A expectativa de sobrevida livre de choques apropriados foi de 74,2 por cento±9,0 após um ano, e de 66,7 por cento±10,7 após três anos. Observou-se diminuição da QV nos aspectos físicos (61,7±28,7), na vitalidade (64,7±19,1), na saúde mental (65,9±22,7) e nos aspectos emocionais (66,7±38,5). Medo e preocupações relacionados ao CDI foram referidos por todos os pacientes. CONCLUSÃO: A despeito da grande eficácia dessa terapêutica, a incidência elevada de choques interferiu na QV e na adaptação ao dispositivo.


OBJECTIVES: To analyze the incidence and causes of ICD therapies in children and young adults and verify their impact on the quality of life (QoL). METHODS: From March/1977 to February/2006, 29 patients (15.7±5.4 years old) were submitted to ICD implants. Aborted cardiac arrest (41.5 percent), sustained ventricular tachycardia (27.6 percent) and primary prophylaxis of sudden cardiac death (30.9 percent) indicated device therapy. The number of therapies was evaluated by interviewing patients and by ICD diagnostic data. The SF-36 questionnaire was used to measure the QoL and the results were compared to healthy population. The expectative of freedom from ICD therapies were estimated by the Kaplan-Meier method. RESULTS: After 2.6±1.8 years follow-up, 8 (27.6 percent) patients received 141 appropriate ICD shocks due to ventricular tachycardia (6) or ventricular fibrillation (2), and 11 (37.9 percent) patients received 152 inappropriate ICD shocks due to supraventricular tachyarrhythmias (8) or oversensing (3). Expectative of freedom from appropriate shocks was 74.2±9.0 percent and 66.7±10.7 percent after one and three years, respectively. Compared to healthy population, QoL decreased in physical function (61.7±28.7), vitality (64.7±19.1), mental health (65.9±22.7) and role-emotional domains (66.7±38.5). All patients referred fear and concern related to ICD use. CONCLUSION: Despite the efficacy of ICD therapies, the high incidence of appropriate and inappropriate shocks interfered in patients' QoL and adaptation to the device.


Assuntos
Adolescente , Adulto , Criança , Pré-Escolar , Feminino , Humanos , Masculino , Arritmias Cardíacas , Desfibriladores Implantáveis/estatística & dados numéricos , Cardioversão Elétrica/estatística & dados numéricos , Qualidade de Vida , Arritmias Cardíacas/etiologia , Arritmias Cardíacas/terapia , Morte Súbita Cardíaca , Parada Cardíaca/complicações , Parada Cardíaca/terapia , Estimativa de Kaplan-Meier , Inquéritos e Questionários , Taquicardia Supraventricular/complicações , Taquicardia Supraventricular/terapia
11.
REBLAMPA Rev. bras. latinoam. marcapasso arritmia ; 20(1): 7-12, jan.-mar. 2007. ilus, tab, graf
Artigo em Português | LILACS | ID: lil-465753

RESUMO

Em pacientes com disfunção ventricular e distúrbio de condução pelo ramo esquerdo, a ressincronização cardíaca é utilizada no tratamento da insuficiência cardíaca (IC) refratária ao tratamento clínico, com resultados comprovadamente eficazes em curto e médio prazos. No final da década de 90, foram introduzidos métodos para cateterização e estimulação com eletrodos específicos para o seio coronário. Atualmente, a técnica endovenosa é considerada o principal método para ressincronização cardíaca; entretanto, a toracotomia ainda é um método efetivo e eficaz nos casos de posições instáveis, altos limiares, estimulação frênica, trombose venosa, entre outros. O presente estudo teve como objetivo descrever a prevalência de toracotomias na casuística do InCor-HCFMUSP desde os primeiros casos (fevereiro de 1997 a dezembro de 1999, quando era a única opção disponível) até dezembro de 2005, listando os principais motivos de sua realização.


Assuntos
Humanos , Insuficiência Cardíaca/terapia
12.
Rev. bras. cir. cardiovasc ; 20(4): 392-397, set.-dez. 2005. graf
Artigo em Português | LILACS | ID: lil-423291

RESUMO

OBJETIVO: Avaliar a evolucão tardia de criancas com marca-passo definitivo por bradicardia pós-operatória, e identificar fatores de risco para a mortalidade. MÉTODO: De 1980 a 2004, 120 criancas receberam implante de marca-passo definitivo por bradicardia pós-operatória. O intervalo médio entre a correcão do defeito e o implante foi de 1,2 n 2,8 anos, com mediana de 21 dias. Bloqueio atrioventricular esteve presente em 94,2 por cento dos pacientes. A via de acesso transvenosa (78,3 por cento) e marca-passos (MP) definitivos ventriculares (79,2 por cento) foram os mais utilizados. Empregou-se o método de Kaplan-Meier e o teste de Log-Rank para a análise de sobrevivência. RESULTADOS: Após 5,7 n 5,9 anos de seguimento (máximo= 22,5 anos), 97 pacientes estavam vivos e 11 haviam sido perdidos para o seguimento. As principais causas de morte foram insuficiência cardíaca (10), infeccão não relacionada ao marca-passo (seis) e morte súbita (três). A expectativa de sobrevida aos cinco, 10 e 15 anos de seguimento mostrou, respectivamente, índices de 80,9 n 4,1 por cento, 75,4 n 5,5 por cento e 67,2 n 7,4 por cento. A persistência de problemas hemodinâmicos após a correcão (correcões paliativas, uso de próteses valvares ou defeitos residuais) foi identificada como única variável preditora independente de risco para mortalidade, alterando significativamente as curvas de sobrevivência (p = 0,0123). CONCLUSÕES: O implante de marca-passo em casos de bradicardia pós-operatória possibilitou boa expectativa de sobrevida. A realizacão de correcões paliativas, assim como a presenca de defeitos residuais ou de próteses valvares, foram os únicos fatores preditores de mau prognóstico para essas criancas.


Assuntos
Recém-Nascido , Criança , Adolescente , Masculino , Feminino , Humanos , Bloqueio Cardíaco/terapia , Estimulação Cardíaca Artificial/efeitos adversos , Seguimentos , Procedimentos Cirúrgicos Cardíacos/efeitos adversos
13.
Rev. bras. cir. cardiovasc ; 20(2): 123-128, abr.-jun. 2005. graf
Artigo em Português | LILACS | ID: lil-413756

RESUMO

OBJETIVO: A estimulação cardíaca permanente melhora a sobrevida de crianças com bradicardia congênita ou adquirida, embora a mortalidade após o implante de marcapasso permaneça relativamente alta. O objetivo deste estudo foi avaliar os resultados em longo prazo de crianças submetidas a implante de marcapasso endocárdico pela veia femoral, incluindo a identificação de fatores de risco associados à mortalidade. MÉTODO: De 1981 a 2000, 99 pacientes variando em idade de um dia a 13 anos (4,1 ± 3,6 com mediana = 3 anos) foram submetidos a implante de marcapasso permanente pela via femoral devido a bradicardia de origem congênita (39,4 por cento), pós-cirúrgica (54,5 por cento) ou adquirida não cirurgicamente (6,1 por cento). RESULTADOS: Ao final de 7,1 ± 5,3 anos (708,3 pacientes-anos) de seguimento, 18 (18,2 por cento) pacientes haviam morrido. A sobrevida atuarial foi de 85 por cento, 79,5 por cento, e 74,2 por cento, aos cinco, 10, e 15 anos, respectivamente. Os fatores independentes de mortalidade identificados pela análise proporcional de Cox foram: menor idade ao implante (p = 0,028), presença de anomalias cardíacas não corrigidas ou presença de próteses intracardíacas (p = 0,0001) e evidências radiográficas de cardiomegalia (p = 0,035). CONCLUSÕES: A estimulação cardíaca endocárdica permanente pela via femoral apresenta expectativa de sobrevida comparável a outras técnicas, com baixas taxas de complicações devidas ao implante de marcapasso. A sobrevida em longo prazo foi limitada pela menor idade e dilatação cardíaca no momento do implante, assim como pela presença de defeitos cardíacos sem correção ou de próteses valvares.


Assuntos
Humanos , Criança , Bradicardia/congênito , Estimulação Cardíaca Artificial , Mortalidade Infantil , Bloqueio Cardíaco , Avaliação de Resultados em Cuidados de Saúde , Pediatria , Fatores de Risco
14.
Pacing Clin Electrophysiol ; 26(1P2): 487-91, 2003 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-12687874

RESUMO

The femoral vein has been used as an alternative conduit to implant pacemakers in children of any weight. Such method associates endocardial pacing and good cosmetics. The aim of this study was to evaluate prospectively, since 1981, the long-term follow-up of 99 children, from newborn to 13 years old (average = 4.1 +/- 3.6 years, 56 girls), who underwent the implantation of pacemakers via the femoral vein. Atrioventricular block was present in 88% of patients, of congenital etiology in 39% and postoperative in 54%. Single chamber pacemakers were implanted in 92% of patients. During a mean follow-up of 5.3 +/- 5.0 years (maximum = 18.2 years), 5 patients died of cardiac causes, 4 of infection, 2 suddenly, and 3 of unknown causes. The 5-, 10- and 15-year actuarial survival rates were 83.7%, 75.7%, and 75.7%, respectively. Transfemoral leads were used for a mean of 48.9 +/- 44.0 months. Reasons for lead explantations were pacing failure in five patients, infection in eight, and elective in nine. The 2-, 5- and 10-year actuarial survivals of transfemoral leads were 87.6%, 73.8%, and 31.8%, respectively. The mean lead survival was 97 months. Overall, 105 reoperations were performed, 38 for battery depletion, 24 for body growth, 14 for infection or pocket revisions, and 27 for miscellaneous reasons. In conclusion, the durability and overall long-term performance of transfemoral leads were excellent.


Assuntos
Marca-Passo Artificial , Adolescente , Criança , Pré-Escolar , Remoção de Dispositivo , Eletrodos Implantados , Feminino , Veia Femoral , Seguimentos , Humanos , Lactente , Recém-Nascido , Masculino , Marca-Passo Artificial/efeitos adversos , Complicações Pós-Operatórias , Análise de Sobrevida
15.
REBLAMPA Rev. bras. latinoam. marcapasso arritmia ; 16(1): 11-16, Jan-Mar. 2003. ilus
Artigo em Português | LILACS | ID: lil-407498

RESUMO

A remoção de cabos-eletrodos de marcapassos e cardioversores-desfibriladores é um problema técnico de difícil solução, razão pela qual as técnicas usadas no procedimento estão sendo continuamente aperfeiçoadas. A utilização do Excimer LASER representa grande avanço tecnológico, contribuindo decisivamente para o sucesso do procedimento. O objetivo do presente trabalho foi analisar criticamente algumas técnicas utilizadas na retirada de cabos-eletrodos de marcapasso e cardioversor-desfibrilador implantável (CDI) e descrever a experiência inicial do Instituto do Coração com a técnica de extração LASER-assistida (Excimer LASER). Foram operados 36 pacientes, com idade variando de 1,4 a 91 anos, sendo retirados 47 cabos-eletrodos de marcapasso e 9 de desfibrilador. As indicações para a remoção incluíram: 19 pacientes com infecção no sistema, 13 com troca eletiva e 5 com outras causas. Em 54 cabos-eletrodos removidos, obteve-se 96,4 por cento de sucesso. A técnica de remoção com excimer LASER apresentou resultados excelentes, equivalentes aos descritos na literatura


Assuntos
Humanos , Pessoa de Meia-Idade , Relógios Biológicos , Terapia a Laser , Desfibriladores Implantáveis , Estimulação Cardíaca Artificial , Terapia com Luz de Baixa Intensidade/tendências
16.
REBLAMPA Rev. bras. latinoam. marcapasso arritmia ; 16(1): 43-48, Jan-Mar. 2003. ilus
Artigo em Português | LILACS | ID: lil-407502

RESUMO

Os autores relatam o caso de uma criança de 23 meses de idade, portadora de síndrome do QT longo congênito,submetida a implante de cardioversor-desfibrilador atrioventricular. A criança apresentou três episódios de fibrilação ventricular revertida após manobras convencionais de ressuscitação cardiopulmonar e desfibrilação transtorácica externa, previamente ao implante. Devido a descontinuidade da fabricação de placas epimiocárdicas, foi utilizado a via transtorácica para o implante dos cabos-eletrodos, com abordagem epimiocárdica para o eletrodo atrial e endocárdica transatrial para o implante do eletrodo ventricular convencional


Assuntos
Humanos , Masculino , Lactente , Estimulação Cardíaca Artificial , Síndrome do QT Longo/congênito , Síndrome do QT Longo/diagnóstico , Síndrome do QT Longo/fisiopatologia , Síndrome do QT Longo/prevenção & controle , Fibrilação Atrial/fisiopatologia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...